Measuring
Insert Earphone RETSPL using Hearing Aid Couplers
There
are certainly clear advantages for using insert
earphones for obtaining accurate audiological
measurements. Insert
style earphones eliminates the need for headbands (thus
increasing patient comfort),
prevents collapsing ear canals, and allows for hearing
tests under higher ambient noise levels.
The use of insert style earphones allows an
increase in interaural attenuation thus reducing the need
for masking.
Audiologists
may find it interesting to know that there is currently a
disagreement and a certain degree of uncertainty regarding
the RETSPL for insert phones generated in HA couplers.
More specifically, the RETSPL concerning HA coupler
type output SPL variability and correction levels between
HA coupler types.
Variability
in RETSPL values for ER-3A insert phones used with the
HA-1 coupler exists because of differences in physical
dimensions, total volume and sound inlet hole diameter
across several models of the HA-1 coupler.
However, Larson & Cooper (1988) found that HA-1
coupler output SPL variability to be clinically acceptable
for ER-3A measurements.
Currently, the output SPL variability for HA-2
coupler types using insert style earphones are not known.
The
issue of coupler output SPL variability is directly
applicable to Audiologists because these are the couplers
used to calibrate the accuracy of the insert earphones.
Using a hearing aid coupler allows measurements of the SPL output
response for the insert earphones.
A
study by Frank et al. 1991 examined three hearing aid
couplers (HA-1, HA-2 and DB-0138 (HA-2 with entrance
through a rigid tube)) with two pairs of insert earphones
(ER-3A left and right, EARTONE-3A left and right).
For each coupler types, the mean coupler SPLs were
found to be very similar across the four insert earphones,
which is not surprising since both the EA-3A and EARTONE
insert phones are manufactured to identical
specifications.
It
is important to note that it is not uncommon to find
output variations of 2 to 3 dB within the same earphone
type.
For example, at 250 Hz, the ER-3A (R) and at 8000
Hz ER-3A (L) demonstrated lower than average coupler SPLs
compared with other insert phones. This trend was visible across all couplers types in the Frank
et al. 1991 study.
The
HA-1 and HA-2 coupler SPLs are very similar for each
insert phone within a range of 125-8000 Hz.
The response characteristic in the range of 125 -
3000 Hz was generally flat.
The DB0138 coupler has a similar response as above
except that the flat frequency response occurs between 125
– 4000 Hz.
The
input frequency for each insert earphone has a +/- 1 Hz
tolerance. In
clinical practice, audiometers have a +/- 1% frequency
tolerance for type 1 audiometers, +/- 2% frequency
tolerance for type 2 and +/- 3% for type 3, 4 and 5 (ANSI
S3.6 – 1996). What
Frank et al. 1991 was concerned with was the steep high
frequency response characteristic for each earphone.
Audiologists should know that there is a general
rule that measurement repeatability decreases if any
transducer that is operated at a discrete a frequency that
lies along a steep slope (click here
for a related related article)
The
following is a review of the redsults and findings of
Frank et al. 1991.
The
SPL variations were less than 1 dB for both the HA-1 and
HA-2 coupler between the range of 125 – 3000 Hz and 125
– 4000 Hz for the DB0138.
The
SPL variations were greater than 1 dB for the HA-1 and
HA-2 SPL between the range of 4000 – 8000 Hz with a
maximum variation of 4.1 dB at 8000 Hz.
For the DB0138 this range is 6000 – 8000 Hz with
a maximum variation of 4.6 dB occurring at 8000 Hz.
Higher
frequency output variations can be directly related to the
steeply sloping high frequency response for each earphone
measured in each coupler.
For
specifying high frequency insert earphone RETSPL, a +/- 3%
frequency tolerance should be avoided (especially at 8000
Hz), hence type 3 audiometers and above should be avoided,
unless they can be provide frequency accuracy in the +/-
1% range. Clinically
it is recommended that the audiometer frequency should be
as close as possible to the nominal center frequency so
that insert earphone output variations that could
influence calibration and testing are minimized.
From
the Frank et al. 1991 study, all three hearing aid
couplers provided stable coupler SPLs.
However, Frank et al 1991 recommends the use of the
DB0138 coupler (HA-2 coupler with entrance though a rigid
tube) to be used for referencing normative thresholds and
for routine clinical calibration of ER-3A and EARTONE 3A
insert earphones. Their
reason for selecting the DB0138 is because it is the
easiest one to use for calibration. It is mentioned that Larson and Cooper (1989) recommend the
HA-2 coupler. Frank
et al. 1991 found that when using the HA-2 coupler,
placement of the eartip tube flush with the earmold
substitute sound channel outlet may be difficult to
obtain because the eartip termination point cannot be
visualized when the HA-2 coupler is intact.
In addition, they found that some eartip tubes do
not fit as tightly as others in the earmold substitute
sound channel once the eartip has been removed, hence
questioning the adequacy of the seal.
Note: This site is
best viewed using IE 4.X or higher under 800x600
resolution
|