Instrument
Specifications for Aural Acoustic Immittance
The
specification for Aural Acoustic Immittance is described
in ANSI S3.39 1987 and revisited and simplified in
this article below. According
to ANSI S3.39 1987,
The purpose of
this standard is to ensure that acoustic-impedance or
acoustic-admittance measurements will be substantially the
same for a given ear when these measurements are obtained
with any instruments that meet the specifications and
tolerances outline in this standard, and when comparable
test conditions prevail.
This standard is not intended to inhibit
development, to restrict incorporation of new features
(such as probe frequencies other than 226 Hz), or to
discourage other improvements that may become useful in
the measurement of aural acoustic impedance or acoustic
admittance.
Acoustic
immittance machines are categorized into specific types
(1-4) according to its features and capabilities, similar to
how audiometers are categorized.
Those devices that
offer the greatest number of features and capabilities are
classified as Type 1 and the most basic devices are
categorized as Type 4 (see
chart). A
type 1 acoustic immittance device generally is more precise than a type 2 or 3.
There are however, mandatory characteristics and features
acoustic
immittance devices must have in order to meet ANSI S3.39
standards.
Ambient temperature and barometric
pressure are important variables which affect the
measurement of acoustic immittance. Since these
variables are likely to change on a periodic basis due to
environmental conditions, it is important that
Audiologist's calibrate their machines for changes in
these conditions, on a daily basis prior to use.
According to ANSI S3.39 -1987, the manufacturer will
provide the hard walled test cavities and data necessary for this
calibration. There should be at a minimum, three
calibration cavities provided with Type 1-3 devices with
volumes of 0.5, 2.0 and 5.0 cm3.
Usually the daily
calibration procedure simply requires
the Audiologist insert the probe end of the immittance
device into the correct sized test cavity and press a
calibrate button, however each manufacturer may adapt
different methods of calibration. It is up to the
manufacturer to provide the recommended procedure for
using the calibration cavities.
In addition, each manufacturer must also specify the temperature and
barometric pressure range in which the immittance device
will be able to maintain optimum performance.
ANSI S3.39-1987 requires that all devices be able to
operate within a range of + 15 oC to + 35 oC,
over a relative-humidity range of 10% - 90%.
Type
1, 2 and 3 devices must be able to change and indicate the air
pressure in the external auditory meatus and in the
manufacturer calibration cavities.
The maximum amount of pressure permissible by any
immittance device is
- 800 daPa and + 600 daPa.
The minimum pressure requirements is dependent on
the type of device. A
type 1 and 2 immittance device is minimally required to
operate in the range of 600 daPa to + 200 daPa.
A type 3 devices minimum range is 300 daPa to
+100 daPa.
It is
the manufacturers responsibility to indicate the accuracy
and limitations of their device, as well as the
limitations with regards to the atmospheric pressure and
altitude above sea level in which the device will operate. ANSI S3.37 1987 indicates that type 1 and 2 devices must
meet an accuracy of +/- 10 daPa or +/- 10% of the reading
on the display (which ever one is greater).
Type 3 and 4 devices must be accurate to +/- 10
daPa or +/-15% (which ever is greater).
All
immittance devices regardless of type, must have a 226 Hz
probe signal with an accuracy within +/- 3% of that signal
with no more than 5%
total
harmonic distortion.
Other signals may be present within a device, but
only 226 Hz is required for compliance with standard.
The
total harmonic distortion level allowable for frequencies
other than the 226 Hz probe tone differs depending on the
transducer being used. A
chart summarizing the maximum total harmonic
distortion for pure-tone acoustic
reflex activating signals can be found by clicking
here.
For signals that fall at higher level settings, the
max total harmonic distortion is not to exceed 5% for
supra aural earphones or 10% or insert style earphones.
The fixed frequency reflex activating stimuli on
any immittance device should be any number of the
following: 250,
500, 1000, 1500, 2000, 3000, 4000, 6000, or 8000 Hz.
Type 1 devices must have the following; 500, 1000,
2000, and 4000 Hz for both contralateral
and ipsilateral
acoustic reflex measurements.
Note: This site is
best viewed using IE 4.X or higher under 800x600
resolution
|