|
Calibrating
Bone Conductors using OAE’s
The
most common method of calibrating an audiometer for bone
conduction testing requires an artificial mastoid where
the bone transducer is positioned on the mastoid and an
electrical input signal is measured and equated into a
certain sound pressure level.
The
problem with using an artificial mastoid is that it does
not represent the anatomical uniqueness of each person who
will be tested.
That is, a bone conductor calibrated using an
artificial mastoid is calibrated to replicate the
mechanical impedance of the head.
But not all heads are the same.
A static force of approximately 5.4 newtons (N)
(ANSI, 1996) is the current standard, but this varies
depending on the size of a person’s head.
In
this article, an objective method of
calibrating a bone conductor using otoacoustic emissions (OAE’s)
will be discussed.
Distortion product OAE’s are elicited using two
pure tones (f1 and f2).
Most normal ears will produce a recordable emission
at 2f1 – f2, where the magnitude of the emission varies
as the magnitude of f1 (L2) and f2 (L2) are varied.
The
theory underlying this calibration technique is that
OAE’s are not restricted to pure tone air conducted
stimuli, but can also be elicited using pure tone bone
conduction stimulus.
Barany (1938), Lowy (1942), Tonndorf (1966) have
shown that the cochlea responds equally to both air and
bone conducted stimuli.
That is, a traveling wave will be established along
the basilar membrane regardless of whether it is elicited
by an air or bone conducted stimuli (Zwislocki, 1953).
DPOAE
with stimulus magnitude can be represented using a contour
plot that is illustrated below.
In the graph below, L1 and L2 (level/magnitude of f1
and f2 tone in dB SPL) are represented on the Y
and X-axis respectively.
The lighter shaded region represents the space
where the highest magnitude DPOAE’s were recorded (Ldp
= 5-10 dB SPL).
The darkest regions represent the area where the
DPOAE’s were weakest (Ldp = -20-15 dB SPL).
At these levels, it is difficult to distinguish
whether the response is a DPOAE or whether it is simply
part of the noise floor.
figure
1 from Purcell et 1999
If
we were to hold either L1 or L2
constant in magnitude, then a graph with a single curve
can be obtained.
This graph is called an IOgram and is illustrated
below with L1 held constant.
figure
2
from Purcell et al 1999
The
IOgram is used to indicate nonlinear emission growth with
changes in L2.
That is, with L1 held constant, we can
see how the magnitude of the emission changes as the
magnitude of the f2 tone (L2) is
varied.
A general pattern with DPOAE’s as seen on an
IOgram is that emissions tend to increase up to a certain
level, and then level off.
This very pattern is seen in the above IOgram where
the growth decay is quite obvious.
According to Purcell et al.
1999, it is advisable to hold L1
constant and vary L2 rather than the reverse
because, “emissions tend to be recordable above the NF
(noise floor) for a larger range of L2 and of L1”.
So
what do these graphs mean towards calibration you may ask?
Well, the IOgrams are used to compare emissions
elicited through BC with those from AC.
As I mentioned before, a critical point about
DPOAE’s is that they can be elicited by either BC or AC
because both methods set the basilar membrane into motion.
Using
IOgrams with a fix L1 value ensures that
emissions are compared in the same region in stimulus
space.
Once
again, let’s review these important steps.
1.
f1 and f2 tones are fixed at a certain frequency
2.
The L1 magnitude is fixed and L2 is free
to vary in order to find the maximum emission level
3.
Two IOgrams will then be generated, one by AC and
the other by BC
It
is critically important that L1 remains constant during
this procedure, therefore it is recommended that the same
stimulus transducer be used during all measurements.
Purcell et al. 1999 recommends that f1
is always presented using an air speaker in a probe
inserted in the ear canal.
This same probe contains the microphone for the
response measurement as well as a second speaker for the f2
tone when performing an AC measurement.
When a BC measurement is being performed,
everything remains the same, except that instead of using
the probe to generate the f2 tone, a bone
conductor at f2 is used.
According
to Purcell et al. 1999, if the IOgrams are similar under bone
and air conduction stimulation, it follows then that the f2
tone must be reaching the cochlea with similar magnitudes
in both cases.
For
example, if the f2 tone by air conduction produces an
IOgram that looks like figure 2 above where the maximum
emission occurs at 50 dB SPL, we can expect that the same
f2 tone by bone conduction will produce a similarly
looking IOgram.
If the maximum emission by bone conduction does not
occur at 50 dB SPL or close, then it is safe to assume
that the bone conductor is not properly calibrated.
When this happens, the maximum emission point by BC
must be located and marked as being 50 dB SPL because this
is the point of maximum emission by AC (assuming that AC
is properly calibrated).
By repeating the IOgram correlations through the
frequency range of interest, the bone conductor can be
objectively calibrated against the known air stimulus
(Purcell et al. 1999 pg 378).
Note: This site is
best viewed using IE 4.X or higher under 800x600
resolution
|