Reference
Equivalent Threshold Force Levels For bone Vibrators
Bone
vibrators used during audiometric evaluation are
calibrated using an artificial mastoid, as specified by
ANSI 1987 guidelines. During calibration, the
vibrator is coupled to the mechanical coupler with a
static force of 5.4 +/- 0.5 N and the artificial mastoid
is equipped with an electromechanical transducer that
measures the force level at the point where the vibrator
is in contact with the surface of the artificial
mastoid. A Bruel & Kjaer artificial mastoid type
4930 is generally used for the calibration of bone
vibrators. The typical setup is diagramed
below.
An
artificial mastoid needs to exert an impedance onto the
tip of the bone vibrator that will have an impedance which
is similar to an average human head-bone situation
over the range of test frequencies (normally 250 – 4000
Hz). Artificial
mastoids that meet ANSI S3.26 1981 standards must not
differ from the manufacturer’s values by more than +/-
25% over the range of 250 –4000 Hz.
If this condition is not met, the artificial
mastoid must be re-calibrated as indicated in the
manufacturer’s manual.
Calibration should be done under an ambient room
temperature of 23 +/- 2oC.
Every artificial mastoid, regardless of who
manufactured it, needs to include a manual with a curve or
table indicating the mechanical impedance vs. frequency
characteristics of the device.
Reference
equivalent threshold force level calibration values for
the B&K type 4930 (with serial
numbers higher than 526226) used with a Radioear B-71
attached to a P3333 headband are indicated below.
|
Frequency
(Hz)
|
rms
force level
|
|
250
|
61
|
|
500
|
59
|
|
1000
|
39
|
|
2000
|
32.5
|
|
3000
|
28
|
|
4000
|
31
|
Values taken from ANSI S3.26 1981
If
another artificial mastoid and bone vibrator combination
is to be used, calibration values can be determined
using subjective
techniques.
Audiologists
need to know that when the bone vibrator is placed across
the front of the head, rather than on the mastoid, more
force needs to be added to obtain thresholds comparable to
the mastoid placement.
With a mastoid placement, the recommended level of
force against the mastoid is 5.4 +/- 0.5 N.
The following values are the correction factors
that must be added to the reference equivalent threshold
force level calibration figures above for forehead
placement of the bone vibrator.
|
Frequency
(Hz)
|
Correction
Value (dB)
|
|
250
|
13.5
|
|
500
|
15.0
|
|
750
|
12.5
|
|
1000
|
10.0
|
|
1500
|
9.0
|
|
2000
|
8.5
|
|
3000
|
7.5
|
|
4000
|
6.5
|
Note: This site is
best viewed using IE 4.X or higher under 800x600
resolution
|