Calibration
of Speech Signals
ANSI
Requirements for Speech Audiometers (ANSI S3.1 section 5.2
)
When
speech test material is used, the presentation level must
meet a standard reference level according to ANSI S3.1.
This reference level can be monitored using a VU
meter. The
speech audiometer VU meter pointer should reach 99% of the
input signal value in 300 msec and then overshoot by 1 to
1.5%. The
best way to check this is to use a calibration tape
recording, which contains a 1000 Hz tone.
With this tone, the Audiologist can visually
monitor on the VU meter that the pointer or needle
accurately points to 0 dB.
Sound
Pressure Level of Speech (ANSI S3.1 section 5.3)
According
to ANSI, the sound pressure level of a speech signal at
the earphone is defined as the rms sound pressure level of
a 1000 Hz signal adjusted so the at the VU meter
deflection of this signal is equal to the average peak VU
meter deflection produced by the speech signal.
Usually, pre-recorded speech signals and the 1000
Hz calibration tone will be at the same level. If however they are not, then the VU meter sensitivity can be
adjusted as needed during testing.
Standard
Reference Threshold Sound Pressure Level for Speech (ANSI
S3.1 section 5.4)
The
standard reference threshold SPL for speech through an
earphone is 10 dB re 20 u Pa.
What this means is that on the audiometer, a
reading of 0 dB HL will correspond to 19 dB SPL.
This figure was attained using a Western Electric,
type 705-A earphone on a NBS
9-A coupler. When
referenced to other earphones, the proper correction
factors must be used.
The correction factors for other popular earphone
manufacturers are listed in table 1.

Reference-level
calibration can be done by setting the 1000 Hz signal at 0
V through the microphone and measuring the output SPL of
the 1000 Hz signal at a HL dial setting of 60 dB.
ANSI standards allow a 3 dB tolerance around this
value. Therefore,
according to the above table, a TDH-39 earphone can be
considered within acceptable calibration range if it falls
within 77 – 83 dB SPL (80 +/- 3dB).
The
calibration equipment used for the reference-level
calibration can be the same as that used for SPL
measurements with tones.
The 1000 Hz reference signal needs to be from the
same source as that for speech audiometry. This can either be a microphone (live voice), or prerecorded
material (tape or CD).
If you
chose to use a microphone, which most Audiologists tend to
use, the external calibration tone needs to be applied in
an anechoic space. The
reason for this is to ensure that the measurement of the
audiometer characteristics is independent of the acoustic
environment in which the microphone is placed.
In order
to calibrate the microphone in an anechoic condition, you
may place the microphone in a hearing aid test box (FONIX
BOX or AUDIOSCAN). The
advantage of this approach is that the control microphone
of the test box can ensure that the level of the test tone
at the audiometer mic is correct.
Another means of calibration the audiometer mic is
to use a portable audiometer or external oscillator
connected to an earphone or preferably a speaker can be
used to deliver the calibration tone.
When this approach is used the mic and the
transducer should be at the same distance that the talker
will be from it when used in the clinic.
The disadvantage of this method of calibration is
that the acoustic environment (test room) can influence
the measurements taken.
On
two-channel audiometer, you can set up pure tone testing
on one channel and speech testing on the other channel.
What you can do is to use one channel present the
calibration tone to calibrate the other channel.
Then you can reverse this and calibrate the other
channel in the same fashion.
To do this, you can set one channel to live voice
testing and the other channel to present a continuous 1000
Hz tone.
Acoustic
Fidelity (ANSI S3.1 Section 5.6)
Live
Voice Audiometer (ANSI S3.1 Section 5.6.1)
The
frequency response characteristic of a live-voice
audiometer shall be such that with the microphone in an
acoustic sine-wave field of a given sound pressure level
(approximately 74 dB), and in the relationship to the
acoustic source which is recommended by the manufacturer
as to distance and orientation, the sound pressure level
developed by the audiometer at each of the frequencies
200, 300, 400, 700, 1500, 2000, 3000, 4000 Hz does not
differ from that at 1000 Hz by more than +/- 5 dB.
The above statement holds a lot of information.
Basically, the above condition can be met by
following the steps from section 5.4.
That is, instead of using just a 1000 Hz tone to
calibrate your audiometer, you use all the above
frequencies (200 - 4000 Hz), and then determine if it is
within +/- 5 dB of the 1000 Hz tone.
To make sure that the input at each frequency is
close to 74 dB, a monitor microphone should be placed near the
audiometer test microphone.
Recorded
Speech Audiometer (ANSI S3.1 Section 5.6.2)
The
frequency response of a recorded speech audiometer through
the frequency range of 200 – 4000 Hz, shall not differ
from that at 1000 Hz by more than +/- 5 dB.
Any frequencies above 4000 or below 200 Hz cannot
be greater than 10 dB above the values between 200- 4000
Hz. In order
to ensure this, the measurement device must be set to
measure different frequency regions.
Such a device would include a sound level meter
with one octave or one-third octave filters.
According
to the standard, speech audiometers used with recorded
materials must have materials that comply with the Record
Industry Association of America Standards, “Standard
Recording and Reproducing Characteristics and Dimensional
Standards – Disc Phonograph Records for Home Use,”
Bulletin No. E-4. The
problem with this stipulation is that it is extremely
difficult to determine if prerecorded speech material on
discs complies with the standard, especially since
ANSI-1969; R-1973 does not include a standard for recorded
materials. The
recommended approach to this situation is to use tapes,
which contain calibration tones at the appropriate
frequencies.
Overall
Distortion (ANSI S3.1 Section 5.6.4)
Speech
audiometers, like pure tone audiometers must be within tolerable limits of overall distortion present in the
circuit of the device.
Distortion measurements should be made
independently for each speech source (microphone, taped
materials). The
input signal used to measure distortion through a speech
audiometer must be free of any distortion.
According to ANSI this means that the pure-tone
input must have no harmonic greater than 40 dB below the
fundamental.
Distortion
measurements are measured at very high levels, i.e. 120 dB. When
measuring the overall distortion through a speech
audiometer, ANSI requires the fundamental frequency be at
least 25 dB or more above the level of any higher
harmonic.
Pre-recorded
discs or tapes can be used to test for distortion.
When testing the sound path through the microphone,
a pure tone signal with a 74 dB SPL input is used.
A sound level meter with one or one-third octave
filters with filter roll-off greater than 25 dB per octave
can be used to make the distortion measurements.
In order to do so, one must adjust the HL dial so
that each tone (200, 400, 700, 1000, 2000 and 4000 Hz or
250, 500, 1000, 2000 and 4000 Hz) is produced at 120 dB SPL. After this, read the levels at each of the harmonic intervals
above the test frequency to make sure that they fall at or
below 95 dB SPL (120 – 25).
Sound
Pressure Levels for Speech Audiometers (ANSI S3.1 Section
5.7).
Range
and Intervals of Hearing Threshold Levels for Speech (ANSI
S3.1 Section 5.7.1)
Accuracy
of Sound Pressure Level (ANSI S3.1 5.7.2)
On the
physical audiometer, the dial readings should be able to
span a range of 0 – 100 dB in steps of 2.5 dB or less.
The measured difference between to successive
intervals of the audiometer dial and the actual output
should not differ by more than 1 dB.
The SPL
produced by the earphones should not differ from the
audiometer’s indicated value by more than 3 dB. This measurement is made at 1000 Hz at 60 dB.
Since
most audiometers use the same attenuator dial for both
pure tone and speech signals, it is not necessary to
recheck the attenuator for speech signals if the
calibration for pure-tones has been conducted.
Noise
(ANSI S3.1 Section 5.8)
The
noise floor according to ANSI must be at least 50 dB below
the level of the signal.
To make this measurement, the presence and absence
of a 1000 Hz signal is used.
On the audiometer, the gain control is set so that
the VU meter indicates a reference level when the input is
1000 Hz. This
input can be a signal from recorded material or a signal
through a microphone, in which case the signal needs to be
at 85 dB SPL. For
the microphone signal path, the level is set to 85 dB SPL.
The output of the audiometer is measured with the
above input signals set to 100 dB on the audiometer dial.
In addition, the output will also be measured with
no signal input to the audiometer. Signal to noise ratios are then determined by comparing the
output levels with the input and with no input.
Note: This site is
best viewed using IE 4.X or higher under 800x600
resolution
|