Calibrating Audiometers

Audiometer Power Supply Requirements

ANSI S3.6 specifies a standard for the main power supply which is referred to as the "Mains operation".  Mains operation refers to the power source for the audiometer which can be in the form of a supply voltage, battery power or combination of the two.  

For short term voltage reductions, the mains supply can be reduced by up to 100% for a time that is no more than 10 ms, 50% for 20 ms or 20% for 50 ms, as long as the next drop in power is 10 s or more.  

If a longer periods of power fluctuations occur, audiometers are required by ANSI S3.6 section 5.3.1 to operate within manufacturer specifications within a +/- 10% supply voltage variation or +/- 5% if battery power or a combination of battery power is being used.    

If a battery powered audiometer is being used, the limits of battery voltages within the manufacturer specifications needs to be provided.  According to ANSI 3.6 section 5.3.3, audiometers that operate on batteries needs to included a voltage indicator to ensure that the device is operating within the allowable limits.      

            

 

Note:  This site is best viewed using IE 4.X or higher under 800x600 resolution


Michael S.H. Der 
Copyright © 1999 Homeboy Homework Co. 
All rights reserved. Revised: April 13, 2001   

click here to email me     

line   line


The current time is EST

 

Audiometer Topics

Bone Conduction Topics

Sound Field Topics

Immittance Topics

Other Topics


 

Audiometers

Bone Conduction

Immittance

Miscellaneous

 


The University of Western Ontario

School of Communication Sciences and Disorders

 

 

line

Disclaimer | Contact Us | FAQ | About Me! | Links

 

TOTAL NUMBER OF VISITORS