Calibrating
Audiometers
Audiometer Power Supply
Requirements
ANSI S3.6 specifies a standard for
the main power supply which is referred to as the
"Mains operation". Mains operation refers
to the power source for the audiometer which can be in the
form of a supply voltage, battery power or combination of
the two.
For short term voltage reductions,
the mains supply can be reduced by up to 100% for a time
that is no more than 10 ms, 50% for 20 ms or 20% for 50
ms, as long as the next drop in power is 10 s or more.
If a longer periods of power
fluctuations occur, audiometers are required by ANSI S3.6
section 5.3.1 to operate within manufacturer
specifications within a +/- 10% supply voltage variation
or +/- 5% if battery power or a combination of battery
power is being used.
If a battery powered audiometer is
being used, the limits of battery voltages within the
manufacturer specifications needs to be provided.
According to ANSI 3.6 section 5.3.3, audiometers that
operate on batteries needs to included a voltage indicator
to ensure that the device is operating within the
allowable limits.