General
Transducer Precautions
The most
likely earphone types encountered with clinical
audiometers are the Telephonics TDH types and ER-3A insert
phones. There are three common TDH style earphones,
all of which physically look the same. The TDH 39,
49 and 50 along with their audiocups which may be type 51
or type MX-41/AR all look identical, but each have
different frequency response characteristics.
A useful
difference to know regarding the different TDH style
earphones is their impedance values. The TDH 49 only
comes with a 10 ohm impedance, whereas the TDH 39 and 50
come with a variety of different impedance values (10, 60,
300 ohm etc). In general, transducers with a higher
ohm rating are more resistant to noise than low impedance
devices. Noise can occur if the metal contacts of
the transducers become oxidized. When this happens,
any turning or twisting at the point of contact may cause
noise in the form of static or crackling. Oxidation
on the metal contacts can have up to a 15 dB effect on
measured threshold values as found by Rick Cameron of EMI
Canada (personal communication), who was kind enough to
allow me to observe a day with him during the calibration
of the these devices. The
problem of oxidation can easily be solved by using an oil-based solvent cleaner on the transducer
contacts.
In clinical
testing, the only advantage of using lower resistance
transducers is that they do not require as much voltage to
drive the device. However, when doing tests such as
ABR's, lower resistance transducers are often
useful. The reason for this is that the high voltage
needed to drive the high resistance transducers also tend
to introduce more noise into the testing situation.
In threshold testing, this increased system noise is
negligible but in ABR testing, it is critical that as much
system noise is eliminated in order to get the most
accurate measurements, thus the use of lower impedance transducers.
Some
audiometer manufacturers require a transducer of a
specified resistance. For example, GSI instruments
use 50 ohm transducers while Madsen and Interacoustics use
10 ohm. It is important to label transducers with
their impedance values so that the correct earphones get
used on the correct audiometer. In busy hospital
settings, there may be two or three audiometers present in
the clinic, and it is not uncommon for clinicians to
switch transducers from one machine to the other if one of
them happens to break or get damaged. Switching
transducers among different audiometers is strongly
discouraged because each transducer is calibrated
specifically to one audiometer. A study by Valente
et al. (1994) captured the error and variability that can
occur by arbitrarily plugging in an insert earphone to an
audiometer which it was not originally calibrated.
For a summary of this study, click
here.
Note: This site is
best viewed using IE 4.X or higher under 800x600
resolution
|