|
Calibration
of the TDH-39 Earphone
Correction factors
and thus accurate calibration of TDH-39 earphone mounted
on MX-41/AR and Auraldome cushions are difficult to
achieve. The
reasoning behind this is because different methods of measuring
the performance of each device produce different results for each
piece of hardware. When psychoacoustic
studies are conducted comparing the two earphone cushion
combinations, both seem to suggest equivalent performance.
That is, threshold data obtained using both the
supra-aural and circumaural cushions implies that
equivalent SPL’s are produced at the tympanic membrane.
Consider the following data obtained from
Ciechanowski et al. (1976).

Ciechanowski et
al. (1976) conducted similar measurements using an
electroacoustic technique, and found a very different
result. What
they found was that less SPL is generated by the TDH-39
using a circumaural cushion on a flat plate coupler
compared to the same earphone mounted in a MX 41/AR in a NBS-9A
coupler.
|

|
NBS-9A
coupler
|
In conducting an
electroacoustic comparison, the TDH-39 earphone was
mounted in the MX-41/AR cushion and the performance was
measured in a NBS-9A coupler illustrated above.
The Auraldome coupled to the TDH-39 earphone was
measured using a flat plate coupler with a 1-inch
microphone.
Consider the data
and graph below. According to this graph, the two earphone cushion
combinations differ significantly.
The largest discrepancy between the two earphones
occurs at 125 Hz. The
puzzling question now is why do the two earphone cushion
combinations perform similarly under psychoacoustic
testing but not under electroacoustic testing?

Audiologists
may find it interesting to know that the physical
construction of the circumaural earphone has a specific
impact on the behaviour of sound transmission.
Generally, the greater volume of air enclosed, the
greater the reduction in SPL in the low frequencies.
This may account for the large discrepancy at 125
Hz in the above graph.
When the TDH-39 MX
41/AR and Auraldome combinations are fixed onto a KEMAR
manikin, different results occur again.
Consider the following graph, which presents a mean
difference between the SPL’s developed in the Zwislocki
coupler mounted in the manikin.
In this set up, the TDH-39 earphone is set in the
MX 41/AR and Auraldome cushions, and in the NBS 9/A
coupler by the same earphone in the MX/41 AR cushion.

What is
clear in the manikin experiment is that the earphone
performance with the MX-41/AR cushion is different from
that measured on the NBS/9A coupler, especially in the
lower frequencies. Some of these differences can be easily accounted for.
The large reduction in SPL at the low frequencies
on the KEMAR manikin is likely due to an acoustic leak
around the earphone.
The greater response at 8000 Hz on the KEMAR
manikin is most likely due to ear canal resonance effects.
At this point, you
may be confused as to which set of results
are most relevant to your practice.
Take the following example into consideration:
The KEMAR manikin
model (containing a Zwislocki coupler) more closely
resembles the human ear than the NBS-9A coupler which is
basically a hard walled cylinder.
Therefore the increased SPL measured at 8000 Hz in
the manikin experiment may very well be more
representative a clinical situation.

According to
Zwislocki 1971, data on the performance of circumaural
earphones on the Zwislocki coupler have shown that the
coupler more accurately represents real ear data.
The drawback however is that no psychoacoustic data
has been reported with circumaural earphones (as of 1976)
and the ear-like coupler in a manikin.

Zwislocki coupler
The problem with
earphone calibration using a manikin is that
psychoacoustic data obtained on a TDH-39 earphone with
MX-41/AR and Auraldome cushions are very similar, however,
electroacoustic responses of the two seem to differ
consistently across cushions and across couplers.
As a result, Audiologists need to be aware that an
accurate method of calibrating supra-aural or circumaural
earphones on a manikin is not yet available.
What was
demonstrated in the Ciechanowski et al 1976 article is that data on the MX-41/AR
and the Auraldome differ across many different testing
conditions, be it on a manikin, on the flat plate/NBS-9A
coupler or on actual people.
Ciechanowski et al 1976 also state that until an accurate
electroacoustic calibration method is available, the best
means of assessment should be a biological
calibration.
Despite
the recommendation by Ciechanowski et al. (1976), the
current ANSI calibration method and standard should
suffice and are the current accepted methods of
calibration.
According
to ANSI S3.6 1996, calibration of the TDH39 supra-aural
earphones needs to be applied to the coupler with a static
force of 4.5 N. The RETSPL values for the TDH39 are
as follows:
| Frequency
(Hz) |
NBS
9A |
| 125 |
45.0 |
| 250 |
25.5 |
| 500 |
11.5 |
| 750 |
8.0 |
| 1000 |
7.0 |
| 1500 |
6.5 |
| 2000 |
9.0 |
| 3000 |
10.0 |
| 4000 |
9.5 |
| 6000 |
15.5 |
| 8000 |
13.0 |
data from ANSI S3.6
1996.
To calibrate the headphones, vary the
audiometer output until it matches the NBS 9A coupler
values at the specified frequency.
Note: This site is
best viewed using IE 4.X or higher under 800x600
resolution
|