Type in key words to search for your topic interest..

 

Frequently Asked Questions

 

On January 10, 2002, an international reader from Australia asked:  

Hi Michael, thanks for an informative website. My questions are in regards to conducting a free field calibration:
 
Should we use dB(A) ie. A-weighted or dB(lin) ie. unweighted when measuring and calibrating the output at the different reference frequencies?
 
Or should we just state the corrections based on linear SPL measurements assuming the people using the equipment will know how to apply the appropriate corrections?
 
This will also affect the choice of SLM used as an octave band filter will not be needed when performing a calibration if A-weighted SPL's are used (although I appreciate that octave band will still be needed for background/ambient level measurements).
 
I was also wondering if there is as yet any standards in regards to free field calibrations. I understand that in the US you call it sound field calibration. Are there any European/ISO standards in this regard? What are your recomendations in terms of how often the free field calibration should be conducted? The current Australian standard requires booth backgrounds tested once every five years for ears-covered audiometry, and as there is no standard for free field, the general practice is also once every five years, although audiometers and SLM's are generally calibrated yearly.
 
I apologise for the long winded nature of this email and thanks again for an informative and useful website.
 
Regards,
 

-----------------------------------------------------------------------------------------------------------------

Michael Der's response,

I apologize for taking so long for this reply.  I've been busy at work and have only just gotten around to your question.  When you mention the dB(lin) scale, I am assuming you are referring to the dB(C) weighted scale.  Are you attempting to measure the ambient room noise in a sound booth? 
 
The A-weighting scale is the most commonly used of the dB scales.  The reponse characteristic of a dB(a) scale most closely matches that of human hearing.  That is, people do not have equal sensitivity to all frequencies of sound across their dynamic range.  Humans are less sensitive to low and veyr high frequencies, and therfore need to increase the SPL at those frequency in order to detect the signal.  For this reason, most measurements are done in reference to dB(a).  Having said this however, it is not clear why you would consider dB(c) scales unless it is for reseach purposes that are not relevant to human perception. 
 
Regarding your question for a standard for calibrationg sound fields.  There are no ANSI or ISO standards on how to calibrate sound fields, so it is not surprising that no Australian standard exists.  For such a standard to exist, sound booths would have to be standardized essentially as a coupler (i.e a 2 cc coupler, 6 cc coupler etc).  An excellent paper which describes how to calibrate your sound field is :
 
Walker G., Dillon H., Byrne D. (1984) Sound field audiometry: recommended stimuli and procedures. Ear & Hearing. 5(1):13-21.
 
howeverthis is not a standard for sound field calibration.  

I have written several articles in my website in the sound fields section, that provides details on how one can calibrate a sound field (none of which are standardized method, cuz one does not exist).  Here is an except from one such article:

Sound field calibration using warble tones

Sound booths used in sound field testing need to be calibrated so that thresholds measurements correspond to those values obtained using earphones.  Audiologist need to be aware that there are no standard RETSPL values used to calibrate the test booths but there are recommended stimuli and procedures used in calibration.  Although these values do allow for good sound field threshold measurements, Audiologists need to be aware that they may also contribute to measurement variability.  

The problem of standing waves in sound field testing can be reduced by using FM warble tones rather than pure tones.  Standardized RETSPL values are not currently available although several studies have presented a recommended procedure and set of stimuli used in calibration and conduction of sound field tests.  Walker et al. (1984) presented what they believe to be a good set of RETSPL for calibrating sound fields. 

Despite published reports of recommended stimuli (Walker et al 1984, Morgan et al 1979), different manufacturers of audiometers produce devices with slightly different FM warble tones.  In addition, sound booths and sound fields found in clinical testing are rarely the same (acoustically and architecturally) across clinics, which contributes to threshold measurement variability (Staab et al 1972). 

 
(My website is down until Jan 16 for maintenance) 
 
I hope this helps. 
 
Please write me back if further clarification is needed. 
 
Regards,
 
Michael Der
 

----------------------------------------------------------------------------------------------

Reader Responded,

Mike,
 
Thanks for your response. I'm sorry I didn't make myself clear in my first email. I am calibrating the output from a screening audiometer in a free field situation.
 
So I am trying to calibrate to 70dB across the frequency spectrum. I have a 3rd octave band type 1 sound analyser and am familiar with its use, weighting scales etc.
 
I have the meter setup in 3rd octave band mode. I have my colleague set the audiometer to read 70dB at 1000Hz. This is our starting point as no weighting correction is necessary. He then adjusts the output from the audiometer until I read as close as possible to 70dB at 1000Hz at the subjects ear position.
 
He then changes the setting on the audiometer to 500Hz, still reading 70dB on the audiometer screen. My question is do I tell him to adjust the level until my meter reads 70dB or 73dB, which takes into account the -3.2dB correction in the A-weighting scale at 500Hz? The same at 250Hz, should I tell him to adjust the level until I read 70dB or 79dB (-8.6dB at 250Hz). Taking into account that I am doing the reading in 3rd octave band mode which are linear/unweighted SPL's at the frequency bands. (I do appreciate that each SLM has its own linear response.)
 
The reason I ask is that if I am to tell him that adjustments are required taking into account corrections for A-weighting then I don't need a SLM with 3rd octave capabilities. I can just use a type 1 integrating SLM and tell my colleague to adjust the output at all his frequencies until my (broadband) SLM reads 70dB(A).
 
Clear as mud?
 
Ultimately it shouldn't matter because if the audiologist using the room has the required knowledge base, he/she should be able to make any corrections deemed necessary according to the manner in which we present our results which would normally be something along the lines of:
 
Frequency Hz    SPL received dB(lin)    SPL adjusted dB(lin)
125                    73.4                            70.1
250                    68.7                            69.9
500                    71.8                            70.2
 
When we present the unweighted SPL at the test frequencies.
 
Or with A-weighted SPL's:
 
Freq                SPL recieved dB(lin)    SPL adjusted dB(lin)    SPL adjusted dB(A)
125                    89.4                            86.0                            69.9
250                    79.9                            78.8                            70.2
.....
etc...
 
So basically I want to know do you recommend the first unweighted SPL method or second weighted SPL method.
 
Regards,
 

----------------------------------------------------------------------------------------------

Michael Der's response in progress....

 

Note:  This site is best viewed using IE 4.X or higher under 800x600 resolution


Michael S.H. Der 
Copyright © 1999 Homeboy Homework Co. 
All rights reserved. Revised: January 31, 2002   

click here to email me     

line   line


The current time is EST

 

Audiometer Topics

Bone Conduction Topics

Sound Field Topics

Immittance Topics

Other Topics


 

Audiometers

Bone Conduction

Immittance

Miscellaneous

 


The University of Western Ontario

School of Communication Sciences and Disorders

 

 

line

Disclaimer | Contact Us | FAQ | About Me! | Links

 

TOTAL NUMBER OF VISITORS